Should I do a slush test on my Clever Logger?
No. A Clever Logger should not be tested using the Strive for 5 slush test.
The slush test described in Strive for 5 is intended for thermometers, and there are several reasons why it is not appropriate for a Clever Logger.
What is the slush test?
The slush test is a simple way of checking the accuracy of a thermometer at approximately 0°C. A mixture of ice and water is prepared and the temperature measuring device is placed into the slurry.
The important point is that Strive for 5 describes this as a test for a thermometer.
A Clever Logger is a temperature data logger, not simply a thermometer.
Why shouldn't I slush test a Clever Logger?
1. Clever Logger is not waterproof
Strive for 5 specifically warns: “Note: This test is only for water-proof devices.”
Clever Logger is not waterproof and should not be immersed in water or an ice slurry.
2. The test is for thermometers, not data loggers
The slush-test procedure in Strive for 5 is specifically described as a method of checking a thermometer.
Clever Logger is a data logger that automatically records and stores a history of temperature readings. The guideline treats data loggers separately from thermometers, so the thermometer slush-test procedure should not automatically be applied to a Clever Logger.
3. Clever Logger samples too slowly for the test
The Strive for 5 slush test is performed over approximately two minutes.
A Clever Logger used for vaccine monitoring normally records at five-minute intervals.
That means the logger may not even take a new scheduled reading during the period in which the test is supposed to be performed. It therefore isn’t a suitable way of reliably checking a Clever Logger.
What should I do instead?
If you need to verify the accuracy of a Clever Logger, follow the manufacturer’s recommended verification or calibration procedure.
Where independent verification is required, the logger can be compared against an appropriate reference instrument or tested using a suitable calibration service.
Ideally, any verification should also be relevant to the temperatures that actually matter for vaccine storage — particularly the +2°C to +8°C range — rather than relying solely on a test at 0°C.
Can I put my Clever Logger in an ice slurry anyway?
No.
Clever Logger is not waterproof, so placing it directly into an ice slurry risks damaging the device.
Putting it inside a waterproof container isn’t a suitable workaround either. Water significantly interferes with the wireless signal, meaning the logger may not be able to communicate normally while submerged.
More importantly, putting a Clever Logger into water or otherwise using it outside its specified environmental conditions will void the warranty.
The bottom line
Do not perform the Strive for 5 slush test on a Clever Logger.
The test is intended for waterproof thermometers. Clever Logger is not waterproof, it is a data logger rather than a thermometer, and its normal five-minute sampling interval does not suit a test conducted over approximately two minutes.
If you need to verify your Clever Logger’s accuracy, use the appropriate manufacturer-recommended verification or calibration method instead.