Should I do an Ice Slush test on my Clever Logger?

No. The Strive for 5 slush test is intended for waterproof thermometers, not Clever Logger data loggers. Here’s what to do instead.

Should I do a slush test on my Clever Logger?

No. A Clever Logger should not be tested using the Strive for 5 slush test.

The slush test described in Strive for 5 is intended for thermometers, and there are several reasons why it is not appropriate for a Clever Logger.

What is the slush test?

The slush test is a simple way of checking the accuracy of a thermometer at approximately 0°C. A mixture of ice and water is prepared and the temperature measuring device is placed into the slurry.

The important point is that Strive for 5 describes this as a test for a thermometer.

A Clever Logger is a temperature data logger, not simply a thermometer.

Why shouldn't I slush test a Clever Logger?

1. Clever Logger is not waterproof

Strive for 5 specifically warns: “Note: This test is only for water-proof devices.”

Clever Logger is not waterproof and should not be immersed in water or an ice slurry.

2. The test is for thermometers, not data loggers

The slush-test procedure in Strive for 5 is specifically described as a method of checking a thermometer.

Clever Logger is a data logger that automatically records and stores a history of temperature readings. The guideline treats data loggers separately from thermometers, so the thermometer slush-test procedure should not automatically be applied to a Clever Logger.

3. Clever Logger samples too slowly for the test

The Strive for 5 slush test is performed over approximately two minutes.

A Clever Logger used for vaccine monitoring normally records at five-minute intervals.

That means the logger may not even take a new scheduled reading during the period in which the test is supposed to be performed. It therefore isn’t a suitable way of reliably checking a Clever Logger.

What should I do instead?

If you need to verify the accuracy of a Clever Logger, follow the manufacturer’s recommended verification or calibration procedure.

Where independent verification is required, the logger can be compared against an appropriate reference instrument or tested using a suitable calibration service.

Ideally, any verification should also be relevant to the temperatures that actually matter for vaccine storage — particularly the +2°C to +8°C range — rather than relying solely on a test at 0°C.

Can I put my Clever Logger in an ice slurry anyway?

No.

Clever Logger is not waterproof, so placing it directly into an ice slurry risks damaging the device.

Putting it inside a waterproof container isn’t a suitable workaround either. Water significantly interferes with the wireless signal, meaning the logger may not be able to communicate normally while submerged.

More importantly, putting a Clever Logger into water or otherwise using it outside its specified environmental conditions will void the warranty.

The bottom line

Do not perform the Strive for 5 slush test on a Clever Logger.

The test is intended for waterproof thermometers. Clever Logger is not waterproof, it is a data logger rather than a thermometer, and its normal five-minute sampling interval does not suit a test conducted over approximately two minutes.

If you need to verify your Clever Logger’s accuracy, use the appropriate manufacturer-recommended verification or calibration method instead.

30-DAY FREE TRIAL

You're going to love Clever Logger

To prove it, we’ll send you a Starter Kit to try out for a whole month.

Set it up, have a play, explore the features – obligation free.

If you love it, just let us know.

If not, just send it back.

Clever Logger temperature logger with external probe

Logger with Dual Temperature Sensors

QUICK SPECS
Model CLD-01
Type Temperature only with Dual Sensors
Temperature Range Internal sensor: -23°C to +60°C
External sensor: -40°C to +80°C
Humidity Range N/A
Battery Type CR2450
Battery Life Replace every 12 months
Accuracy Internal Sensor:
±0.3℃ (0℃ to +60℃)
±0.3℃ to ±0.7℃ (other temperatures)
External Probe:
±0.5℃ (-20℃ to +40℃)
±1℃ (other temperatures)
Offline Memory approx 24 days logging at 5 minute intervals
Clever Logger temperature logger with external probe

Logger with External Probe

QUICK SPECS
Model CLX-01
Type Temperature only with Probe
Temperature Range -40°C to 60°C
Can operate up to 80°C for short periods
Humidity Range N/A
Battery Type CR2450
Battery Life Replace every 12 months
Accuracy ±0.5℃ (-20℃ to +40℃)
±1℃ (other temperatures)
Offline Memory approx 24 days logging at 5 minute intervals